在线客服系统
News

News

Its main business is the production, manufacture and sales of electrical equipment and testing instruments.

News

News HOME >> News

Why Perform 24 Pre-Breakdowns for a New Oil Cup? Uncover the Secret of Electrode Pre-Conditioning

After purchasing the HZJQ-X1 insulating oil dielectric strength tester manufactured by Baoding Huazheng Electric Manufacturing Co., Ltd., many junior laboratory technicians new to high-voltage oil analysis tests often notice a seemingly unusual mandatory operating rule in Item 4 of Section 5 [Precautions] in the manual: “A new oil cup or freshly cleaned oil cup shall undergo 24 breakdown tests before formal testing. When not in use, the oil cup shall be immersed in clean oil.”

Many may wonder: “The instrument and oil cup are brand new, or the cup has just been carefully cleaned and dried with high-purity petroleum ether. Why can’t we fill it with oil and run the test directly to generate a report? Why are as many as 24 idle pre-breakdowns required? Is this just a redundant waste of time?”

This is by no means an arbitrary requirement of the instrument. It is a rigorous golden rule of the “electrode surface conditioning effect”, summarized by the international high-voltage electrical engineering community based on the physics of high-voltage micro-discharge and the surface state of metal electrodes.

We now systematically reveal the underlying physical mechanism behind these 24 pre-breakdowns:

  1. Eliminate microscopic burrs and adsorbed residual gases from machining and wiping Microscopic tips from machining: Even after high-precision turning and mirror polishing, brand-new brass or stainless steel withstand voltage electrodes still have micron-scale irregularities and tiny metallic burrs under a microscope. Microscopic electric field distortion effect: When high voltage is applied, these minute metallic protrusions generate an intensely localized distorted electric field, which easily triggers local field electron emission at a relatively low voltage and causes abnormally low breakdown voltage readings. Desorption of adsorbed gases: Trace residues from silk cloth fibers during electrode cleaning, as well as monomolecular layers of oxygen and nitrogen adsorbed from air on the metal surface, can form micro-bubble ionization sources under high voltage. “Micro-spark plasma polishing” via 24 electric sparks: Fill the cup with qualified standard insulating oil and carry out 24 consecutive high-voltage spark micro-electrolytic discharges. The instantaneous high-temperature arc melts and blunts the most active nano-scale burrs on the metal surface, while fully desorbing and removing trace adsorbed gases. This homogenizes the electric field both microscopically and macroscopically on the electrode surface.

  2. Form a uniform and stable dynamically adsorbed oil film A freshly cleaned metal electrode has a dry, polar surface. Only after repeated high-voltage arc discharges and thermal shocks can insulating oil molecules form a tightly bonded, dense micro oil film with uniform dielectric properties on the metal lattice surface, eliminating abrupt wetting interfaces between the metal and pure oil. Statistical data shows that the coefficient of variation (CV) of test results is extremely high during the first 10–15 breakdowns (often above 20%). After 20–24 breakdowns, breakdown voltage values quickly converge and stabilize at a very steady level.

  3. Standard pre-conditioning operating procedures Fill with qualified conventional transformer oil: Assemble the electrodes in the new or cleaned and dried oil cup, adjust the electrode gap with a standard gauge, then pour in qualified transformer oil to cover the electrodes by at least 10 mm. Set continuous voltage boosting in the [Custom] menu: Set the number of breakdown cycles to 10, run 2–3 successive rounds to complete a total of 24 pre-breakdown conditioning cycles. Drain the conditioning oil and rinse: After conditioning, drain the oil (which contains trace free carbon generated by discharges). Rinse the cup 1–2 times with the oil sample to be tested, then fill with the test sample for official testing.

Follow scientific principles and respect physical conditioning. After completing these 24 electrode passivation and conditioning runs, the oil cup of the HZJQ-X1 reaches its optimal measurement state and delivers high-quality data with excellent repeatability.

Email sales04@bdhuazheng.com Mobile +86 18330222302 WhatsApp +86 18330222302 Wechat +86 18330222302 Wechat +86 18330222302 Back